10 CFR § 35.60 Medical Use of Byproduct Material

§ 35.60 Possession, use, and calibration of instruments used to measure the activity of unsealed byproduct material.

Current through Oct 5, 2026 · Read all of Part 35

⚠ 2 pending changes to Part 35

(a) For direct measurements performed in accordance with § 35.63, a licensee shall possess and use instrumentation to measure the activity of unsealed byproduct material before it is administered to each patient or human research subject.

(b) A licensee shall calibrate the instrumentation required in paragraph (a) of this section in accordance with nationally recognized standards or the manufacturer's instructions.

(c) A licensee shall retain a record of each instrument calibration required by this section in accordance with § 35.2060.

← § 35.59 Recentness of training.§ 35.61 Calibration of survey instruments. →